Configuration
Auto. interval Time
Do you want the analyses to be carried out at a time interval?
Select [Yes] or [No] with [
◀
] and [
▶
]
confirm with [OK].
[Yes] Analyses are carried out at an automatic time interval.
[No] Do not perform analyses in the automatic time
interval.
Auto. interval Time
Set the Auto. Interval Time at which the analyses are to be
carried out.
You will see an input keyboard on the right.
Now use the arrow keys [
◀
], [
▶
], [
▲
] and [
▼
] to move the
black cursor to the desired digit and press the [OK] key. The
selected digit appears in a frame on the left. Repeat the entry
until the desired analysis interval time appears in the frame.
Now move the cursor to the OK field on the keyboard and
press the [OK] key.
You will see the following information on the left:
Current:
currently programmed value
Minimum:
smallest adjustable value
Maximum:
largest adjustable value
If the value at "Current" corresponds to your wish, you do not
need to enter the number again and can immediately move
the cursor to the OK field in the keyboard and confirm with
the [OK] key.
Control measure
How often should a limit value violation be verified before
reporting?
Press the [OK] key.
Control measure
Use [
▲
] and [
▼
] to set the number of control measurements
to be carried out before a limit value is exceeded is reported.
These are carried out at 3-minute intervals after a limit value
has been exceeded in order to avoid false alarms due to the
counter ion effect of the softening system.
Make your selection and confirm with the [OK] key.