Configuration
For water meter selection
Enter the litres/pulse of your water meter.
You will see an input keyboard on the right. Now use the
arrow keys [
◀
], [
▶
], [
▲
] and [
▼
] to move the black cursor to
the desired digit and press the [OK] key. The selected digit
appears in a frame on the left.
Repeat the entry until the desired number appears in the
frame.
Now move the cursor to the OK field on the keyboard and
press the [OK] key.
(When using this function, the Sycon must be in automatic
mode).
You will see the following information on the left:
Current:
currently programmed value
Minimum:
smallest adjustable value
Maximum:
largest adjustable value
If the value at "Current" corresponds to your wish, you do not
need to enter the digits again and can immediately move the
cursor to the OK field in the keyboard and confirm with the
[OK] key.
Is the water meter a "reed" or "hall" contact type?
Reed for potential-free water meter
Hall for electronic water meters
Make your selection and confirm with the [OK] key.
Set the amount of water after which an analysis is to take
place.
Press the [OK] key.
Set the unit.
Set the amount of water after which an analysis is to take
place.
You will see the following information on the left:
Current:
currently programmed value
Minimum:
smallest adjustable value
Maximum:
largest adjustable value
If the value at "Current" corresponds to your wish, you do not
need to enter the digits again and can immediately move the
cursor to the OK field in the keyboard and confirm with the
[OK] key.