82 • Chapter 7: Changing GX-3R (Pro) Parameters
Common Parameters
Several parameters are always selectable and do not depend on which sub tab is selected.
Serial No.
To change the current GX-3R (Pro) serial number, place the cursor in the Serial No. field and enter the
desired serial number. If you change the serial number, once you upload the changes to the GX-3R
(Pro) and return to the main program window, the PC Program will indicate that the GX-3R (Pro) with
the old serial number was disconnected. A GX-3R (Pro) with the new serial number will then connect
to the PC Program.
Station ID
The current station ID will display. If you want to select a new station ID:
1.
Go to the
Station & User
tab.
2.
Click
Read
for the station ID.
3.
Go back to the
Parameter
tab. and use the drop-down menu to select the desired station ID.
4.
To change the station ID list, see “Station & User Tab” on page 96.
User ID
The current user ID will display. If you want to select a new user ID:
1.
Go to the
Station & User
tab.
2.
Click
Read
for the user ID.
3.
Go back to the
Parameter
tab. and use the drop-down menu to select the desired user ID.
4.
To change the user ID list, see “Station & User Tab” on page 96.
Cal Interval (Day)
The calibration interval is the length of time between scheduled calibrations. To change the calibration
interval, use the arrows to the right of the current value to increase or decrease the interval. You may
also highlight the current setting and type in the new desired setting. It can be set anywhere between
1
and
365
days. The factory setting is
90
days.
Maintenance
Mode Tab
(cont’d)
Maintenance Password
(pg. 93)
• GX-3R
• GX-3R Pro
• selected
• deselected
selected (8102)
Battery Change (pg. 94)
• GX-3R
• GX-3R Pro
Any date
Table 6: Parameters
Location
Parameter Tab Name
(pg. # of description)
Applicable
Instrument
Available Choices
Factory