90 • Chapter 7: Changing GX-3R (Pro) Parameters
Warning 2 Time (GX-3R Pro Only)
The
Warning 2 Time
is the amount of time will pass between a Man Down detection and the second
preliminary alarm. It can be set in 1 second increments from
10
-
120
seconds. The factory setting is
75
seconds. Alarm Time
Warning 2 Time
Warning 1 Time.
Alarm Time (GX-3R Pro Only)
The
Alarm Time
is the amount of time will pass between a Man Down detection and the main alarm.
It can be set in 1 second increments from
10
-
120
seconds. The factory setting is
90
seconds. Alarm
Time
Warning 2 Time
Warning 1 Time.
User Mode 2 Sub Tab
Bump Reminder
Selected
: The GX-3R (Pro) will give an indication at start up if it is due for bump testing. The type of
indication will depend on the
Bump Test Past Due Act
setting.
Figure 52: Parameter Tab, User Mode 2 Sub Tab