74 • Chapter 6: Changing SDM-3R Parameters for Standalone Operation
Manual Download
ON
(factory setting): GX-3R (Pro) data can be manually downloaded to a USB drive.
OFF
: GX-3R (Pro) data cannot be manually downloaded to a USB drive.
Power Off After Pass
This is the length of time that passes after a successful bump test, calibration, or alarm check before the
program will automatically shut off the GX-3R (Pro). Use the arrows to the right of the current setting
to change the value or highlight the current setting and type in the desired value. It is defined in
seconds with a maximum setting of
99,999
seconds. The factory setting is
15
seconds. If this parameter
is set to
0
seconds, the GX-3R (Pro) will not be turned off after a successful bump test, calibration, or
alarm check.
Auto Power Off
This is the length of time that passes after the last operation is finished before the program will
automatically shut off the GX-3R (Pro). If a successful bump test, calibration, or alarm check is
performed, the program will automatically shut off the GX-3R (Pro) in the time defined in
Power Off
After Pass
regardless of the
Auto Power Off
setting. Use the arrows to the right of the current setting
to change the value or highlight the current setting and type in the desired value. It is defined in
seconds with a maximum setting of
99,999
seconds. The factory setting is
600
seconds.
Password
Turning the password on (factory setting is off) adds password protection to the SDM Window and to
the Configuration Menu if the SDM-3R is used in Standalone Mode. The factory set password is
0000
but a user selected password can be entered in the space to the right of the
ON
selection box.