Chapter 4: Program Setup • 33
Instrument Information Section
Change Parameter
Selected
(factory setting): The
Edit
function is active, allowing you to change parameters in connected
GX-3R (Pro)s. See pg. 78 for information about instrument parameters.
Deselected
: The
Edit
function is not active and GX-3R (Pro) parameters cannot be edited.
Auto Power OFF Time
This is the length of time that will pass after the last operation is finished before the program will
automatically shut off the GX-3R (Pro). It is defined in seconds with a maximum setting of
99,999
seconds. The factory setting is
3600
seconds. If a successful bump test, calibration, or alarm check is
performed, the program will automatically shut off the GX-3R (Pro) in the time defined in
Power Off
After Pass
regardless of the
Auto Power Off
setting.
Power Off After Pass
This is the length of time that will pass after a successful bump test, calibration, or alarm check before
the program will automatically shut off the GX-3R (Pro). It is defined in seconds with a maximum
setting of
99,999
seconds. The factory setting is
15
seconds. If this parameter is set to
0
seconds, the
GX-3R (Pro) will not be turned off after a successful bump test, calibration, or alarm check.
Data Logger Section
Down Load Data Logger
Selected
: Instrument data is automatically downloaded based on the
Down Load Interval
setting.
Deselected
(factory setting): Instrument data is not automatically downloaded. Instrument data can
always be manually downloaded.
Clear After Down Load
Selected
: The GX-3R (Pro)’s data will be cleared after a data download.
Deselected
(factory setting): The GX-3R (Pro)’s data will not be cleared after a data download.
Down Load Interval
The
Down Load Interval
is the length of time, in days, that will pass since the last download before
the PC Program automatically downloads data from an instrument if
Down Load Data Logger
is
selected. If the instrument is connected to the program before the interval has passed, a data download
will not automatically occur. The
Down Load Interval
can be set anywhere from
1
to
60
days. The
factory setting is
1
day.
SDM Section
Initialize Display Order
Selecting
Initialize Display Order
resets the numbers assigned to specific SDM-3Rs and reassigns
the numbers as you connect SDM-3Rs to the PC Program.
1.
Select
Initialize Display Order
.
2.
Click
OK
and then click
Yes
in the window that appears.