Chapter 7: Changing GX-3R (Pro) Parameters • 93
Disp Follower
Selected
: Zero follower menu item appears in User Mode.
Deselected
(factory setting): Zero follower menu item does not appear in User Mode. Zero follower
menu item is always available in Maintenance Mode.
It is not normally necessary to have the zero follower menu item appear in User Mode. Contact RKI
Instruments before turning this setting on.
Restore Def
Performing a default operation in Maintenance Mode returns all parameters to their RKI factory
settings.
Alarm Latching
Selected
(factory setting): The GX-3R Pro remains in alarm until the alarm condition passes
and
POWER MODE is pressed.
Deselected
: The GX-3R Pro automatically resets an alarm when the alarm condition passes.
Auto Fresh Air Adj
Selected
: The GX-3R (Pro) will ask if you want to perform a fresh air adjustment at the end of the
startup sequence.
Deselected
(factory setting): The GX-3R (Pro) does not ask if you want to perform a fresh air
adjustment at the end of the startup sequence.
Zero Adjustment on Demand
Selected
(factory setting): You can manually perform a fresh air adjust in Measuring Mode by pressing
AIR.
Deselected
: You cannot manually perform a fresh air adjust in Measuring Mode.
LCD Contrast (GX-3R Pro Only)
Higher number means darker background. Lower number means lighter background. Available
settings are
1
-
50
.
20
is the factory setting.
Password Protection
Selected
: A password is required for a zero adjust and power off. The password is
0405
.
Deselected
(factory setting): A password is not required for a zero adjust or power off.
Maintenance Password
Selected
(factory setting): A password is required to enter Maintenance Mode. The factory-set
password is
8102
. Enter a password in the field to the right of the
Maintenance Password
selection
box.