89
Technical Reference
Insert CAL
By using the dedicated preamplifi er NH-08, the NA-42 allows highly accurate
calibration comprising the microphone and preamplifi er (Insert CAL).
Insert CAL function
The microphone sensitivity is determined under the assumption that the mi-
crophone is connected to a preamplifi er with infi nitely large input impedance
and input capacitance of 0.
When calibrating the unit with an electrical signal, the microphone sensitiv-
ity minus a representative value for preamplifi er loss is taken as microphone
sensitivity, and only the internal circuitry is calibrated. With Insert CAL on
the other hand, a signal suitable for the microphone sensitivity is supplied
directly to the microphone, and calibration takes into account the actual
preamplifi er loss.
Microphone
Dedicated preamplifier
Cable
NA-42
Signal input
Calibration signal
Calibration pin