10. Use the supplied screwdriver to adjust the sensitivity control to the
specifi ed value.
When using the sound calibrator NC-74, the calibration values are
as follows.
When using the pistonphone NC-72, adjust the control to the output
sound pressure level value shown on the pistonphone.
11. Turn power to the NA-42 and pistonphone NC-72 or sound calibrator
NC-74 off.
12. Remove the microphone
very carefully and slowly
from the cou-
pler.
UC-11
UC-27
UC-34
UC-26
UC-30
UC-31
UC-52
UC-53A
93.8
93.9
94.0
93.9
93.9
94.0
93.8
93.8
Designation
1 inch
1/2 inch
Nominal outer
diameter
Calibration value
(dB)
IEC 942-1988
JIS C 1515-1991
Sound calibrators
Calibration class 0L
SER.NO
FREQUENCY 250Hz
SOUND PRESSURE LEVEL 114.
dB
Output sound pressure indicated on pistonphone