![Rion NA-42 Instruction Manual Download Page 54](http://html1.mh-extra.com/html/rion/na-42/na-42_instruction-manual_1463440054.webp)
44
Measurement
Low-Pass Filter Setting
The low-pass fi lter has a cutoff frequency of 20 kHz.
Setting procedure
Press the 20 kHz key to toggle the 20 kHz low-pass fi lter between ON and
OFF. When the setting is ON, signal components above 20 kHz are sharply
attenuated.
For details on low-pass fi lter characteristics, please refer to "Technical Refer-
ence" (page 95).
FAST
Hz
kHz
FLAT
SOUND LEVEL
MEASURING AMPLIFIER
HPF
LPF
LEVEL RANGE
CAL
20dB/50dB
POWER
EXT FILTER
RESET
L
p/
L
max
/
L
peak
ON
INPUT
NA-42
ADJ
FAST
SENS
20Hz
20kHz
20Hz
20kHz
MENU
A/C
/FLAT
/SLOW
/IMP
(LIGHT)
20 kHz key
"20 kHz" indication
appears when low-pass
filter is set to ON. No
indication is shown when
filter is OFF.