75
Serial Interface
DOF Transfer Sequence
The instantaneous value sound pressure level output updated every 0.1 seconds
stops when <
SUB
> or DOF0<
CR
><
LF
> is input. Normally <
SUB
> should
be used, because the chance for communication errors is lower.
<
SUB
> : control code 1A
H
Note
Because this unit supports only half-duplex commu-
nication, it cannot receive and send data at the same
time. When sending the <
SUB
> or DOF0 command
to the unit, the computer must use timing which
ensures that the command does not collide with the
instantaneous value data that are being output by
the unit.
NA-42
Computer
Continuous output of instantaneous
value every 100 ms (0.1 seconds)
Output terminated
DOF1<
CR
><
LF
>
(Data) EOT<
CR
><
LF
>
(Data) EOT<
CR
><
LF
>
<
SUB
>