Serial Interface
Commands
RMT n1 n2 n3
Select local mode/remote mode
n1 = 0
Local mode
n1 = 1
Remote mode
n2 n3 = 00 to 15, FF
Set ID number. NA-42 specifi ed by ID number receives
the command. FF specifi es all connected NA-42. To send
a setting command to all connected NA-42 units when
using the SC-31M/SC-31S, use "n2n3=FF".
RMT?
Get ID number and remote mode status
Output data format
p1 p2 p3<
EOT
><
CR
><
LF
>
p1, p2 is ID number
p3 = 1
Remote mode
RNG n
Set level range
Value of level range
n=04 40
dB
n=05 50
dB
n=06 60
dB
: :
: :
n=17 170
dB
n=18 180
dB
RNG ?
Get level range setting
Output data format
p<
EOF
><
CR
><
LF
>
p corresponds to n of RNG.