Preparations
2) Calibration with pistonphone NC-72 or sound calibrator NC-74
For this type of calibration, the pistonphone NC-72 or sound calibrator NC-74
is mounted on the microphone and the sensitivity control is adjusted so that
the sound pressure level indication matches the sound pressure level in the
coupler.
Important
Insert and remove the microphone into and
from the pistonphone NC-72 or sound calibra-
tor NC-74 very slowly and carefully, since the
pressure change could otherwise damage the
diaphragm of the microphone.
1. Turn power to the pistonphone NC-72 or sound calibrator NC-74
off.
2. Turn power to the NA-42 on.
3. Press the SENS key on the front panel.