![Rion NA-42 Instruction Manual Download Page 35](http://html1.mh-extra.com/html/rion/na-42/na-42_instruction-manual_1463440035.webp)
25
Preparations
4. Use the setting ( , ) keys to adjust the sensitivity.
The microphone sensitivity is the value given on the calibration chart
accompanying the microphone (see below).
The preamplifi er loss (caused by static capacitance of microphone and
preamplifi er) is as shown in the table below for various combinations
of microphone and preamplifi er.
Example
1/2-inch condenser microphone UC-52 (sensitivity -32.9 dB) and
preamplifi er NH-17A (loss -0.3):
(-32.9) + (-0.3) = -33.2
The setting is -33.2 (-33.20).
CALIBRATION CHART FOR CONDENSER MICROPHONE UC-52
dB
Open Circuit Sensitivity
-32.9
dB
0dB=1Volt/Pa
22.6
mV/Pa
Cartridge Capacitance:
17.6
pF
Leakage Resistance: >
10
12
at Relative Humidity <85%
Conditions of Tests:
Frequency
1000
Hz
Polarization Voltage:
0
V
Barometric
ressur :
765
mmHg
Relative Humidity
50
%
Temperature
25
Frequency Response Characteristics
No.
37564
Date
01-02-03
Signature
0.1
1
10 Hz
RP48
20
50
100
200
500
1k
2k
5k
10k
20k
40k
+5
0
-5
-10
-15
The microphone sensitivity
Calibration chart
P
e
UC-27
UC-32P
UC-30
UC-31
UC-33P
UC-26
UC-28
UC-52
UC-53A
UC-29
NH-04
NH-12
NH-06
NH-01
NH-11
NH-17
NH-17A
NH-05
-0.1
-0.4
-0.3
-0.9
-0.4
-0.3
Microphone
Preamplifier
1 inch
1/2 inch
1/4 inch
Unit: dB