Preparations
5. Press the SENS key on the front panel to return to the measurement
screen.
6. Use the LEVEL RANGE keys to set the level range to "100" if the
NC-74 is used, or to "120" if the NC-72 is used.
7. When a 1/2-inch condenser microphone is used for calibration, mount
the 1/2-inch adapter to the pistonphone NC-72 or sound calibrator
NC-74.
For 1-inch condenser microphones, the adapter is not required.
Note
For 1/4-inch microphones, the pistonphone NC-72
comes with a 1/4-inch adapter. There is no 1/4-inch
adapter for the NC-74.
8. Insert the microphone
very carefully and slowly
all the way into
the coupler.
9. Set the power switch of the sound calibrator or the pistonphone to ON.
FAST
C
NC-74: 100
NC-72: 120
Level range setting
Select FAST with
FAST/SLOW/IMP
key
Select C with
A/C/F LAT
key
1/2 inch adapter NC-74-002
for NC-74
Coupler
Sound calibrator NC-74
Pistonphone NC-72
Coupler
1/2 inch adapter NC-72-S16
for NC-72
1/4 inch adapter NC-72-S06
for NC-72