Preparations
Calibration
Before starting a measurement, the unit must be calibrated.
There are three types of calibration.
1) Electrical calibration with a tone from a built-in signal generator
2) Calibration with a pistonphone or a sound calibrator
3) Insert CAL calibration
After turning on the unit, wait at least 2 minutes to let the circuits stabi-
lize.
1) Electrical calibration
The unit can be calibrated using the built-in signal generator (1 kHz, sine
wave signal).
Microphone sensitivity adjustment
1. Turn the unit on.
2. Press the SENS key.
3. Use the setting ( ,
) keys to set the value for the microphone
(preamplifi er).
SOUND LEVEL
MEASURING AMPLIFIER
HPF
LPF
LEVEL RANGE
CAL
20dB/50dB
POWER
EXT FILTER
RESET
Lp/Lmax
/
Lpeak
ON
INPUT
NA-42
ADJ
FAST
SENS
MENU
A/C
/FLAT
/SLOW
/IMP
(LIGHT)
20Hz
20kHz
20Hz
20kHz
Setting up/down keys
POWER switch
Sensitivity control
CAL key
SENS key