41
Serial Interface
Model name
: KS-93
Sample fluid volume
xxxxx
: The part varies between 1 and 5 digits (depending on
the measurement time).
vv
: "ML" stands for milliliters and "UL" for microliters.
Measurement value
The initial "A" is a number that has the following meaning
0
: Measurement was carried out normally (including
CELL NG).
1
: Over-range has occurred (particle count has exceeded
eight digits).
2
: An error has occurred during measurement.
xxxxxxxx : A numeral string that indicates the particle count. If the
count has less than 8 digits, the string is padded with
leading zeros. If over-range has occurred, the lower eight
digits only are indicated.
Examples
If measurement was carried out normally:
D/KS-93[100ML],000276916,000019176,000002561,000000396,000000008
<EOL>
If the particle count has exceeded 8 digits (over-range):
D/KS-93[100ML],127856847,109519176,100332561,100052396,100000308
<EOL>
If an error has occurred during measurement:
D/KS-93[100ML],200276916,200019176,200002561,200000396,200000008
<EOL>
Note
Check the contents of the error with the measurement
error information message.