42
Serial Interface
Measurement error information message
The measurement error information message indicates information on the error
status of the immediately preceding measurement. The message is output in
response to the measurement error information request message "Q/E".
The following messages are used.
If two or more errors have occurred, the message for the error with the highest
priority is returned.
(Listed in order of priority)
E/TEMP_NG
<EOL>
During the immediately preceding measurement,
the light source temperature was out of range.
E/LASER_NG
<EOL>
During the immediately preceding measurement,
the laser light source was not operating properly
(laser output level drop, etc.).
E/CELL_NG
<EOL>
During the immediately preceding measurement,
there was a problem with the flow path (cell
contamination etc.).
E/LD_UNSTABLE
<EOL>
During the immediately preceding measurement,
the laser output was unstable.
E/
<EOL>
No error has occurred during the immediately
preceding measurement.
_ indicates a space.
In principle, after receiving the measurement data report message, a check for
errors with the measurement error information message should be performed.
Unlike the measurement data information message, the measurement error
information message can be requested as often as desired until the end of the
next measurement.
Note
Measurement data are output also when an error has
occurred, but there might be a problem with data
reliability.
Summary of Contents for KS-93
Page 2: ......
Page 6: ...iv...
Page 10: ...viii...
Page 18: ...xvi...
Page 21: ...3 Outline Block diagram...
Page 43: ...25 Serial Interface...
Page 69: ...51 Specifications Unit mm Dimensional drawings of main unit...
Page 70: ...Specifications 52 Unit mm Dimensional drawings of power supply unit...
Page 71: ......