7. Maintenance
7-6 Adjusting the flow
94 / 191
NOTE
Press the SHIFT key in step 2 while the flow is displayed to view the following information as you
hold down the key.
・
Diag.
: Self-diagnostic function ON/OFF setting
・
Reg.Diag. : Regular self-diagnostic ON/OFF setting
7-6-2 Manually adjusting the flow (FLOW)
Set the automatic flow adjuster function to OFF to allow manual flow adjustment.
1 In maintenance mode, press the
▲
key or
▼
key to
select [2- 6 FLOW], then press the TEST/SET key.
2 Press
the
▲
key or
▼
key to increase or decrease the
pump output so that the flow for [LineA] and [LineB] are
each more than 0.28 L/min.
Pump output increases 0.5 % each time you press the
▲
key.
Pump output decreases 0.5 % each time you press the
▼
key.
3 Press the TEST/SET key.
The pump output is set. The display reverts to the [2- 6 FLOW] display.
NOTE
Press the MODE key in step 3 to cancel the settings and return to the [2- 6 FLOW] display.
Flow automatic
adjustment: OFF
Flow automatic
adjustment: ON