11. Product Specifications
11-2 Sensor specifications
179 / 191
11-2-3 Semiconductor type
Sensor model
SGF
Detection target
gas
Combustible gas
Toxic gas
Detection range
Depends on detection target gas.
Detection method
Pump suction type
Alarm setpoints
Depends on detection target gas.
Alarm accuracy
(for identical
conditions)
Within ±25 %
Within ±30 %
Alarm delay time
(for identical
conditions)
Within 30 seconds
Within 60 seconds
Alarm type
Two-stage alarm (H-HH)
Operating
temperature range
-10 - 40 °C (no sudden changes)
Operating humidity
range
20 – 90 % RH (No condensation) (moisture content 37 °C 95 % RH or less)
NOTE
Sensor characteristics may degrade significantly in the presence of silicon gases (organic silicon
gases), the various types of halogen gases, sulfide gases, or acidic gases.
Select the calibration gas after accounting for typical humidity in the operating environment.
When calibrating, allow a warmup time depending on the unpowered period in clean air before
proceeding with the zero calibration and span adjustment. Refer to <Warmup time for
semiconductor type sensors> below for warmup times.
Readings may be affected by combustible gases other than the detection target gas.
Readings may fluctuate in the presence of fluctuations (e.g., temperature and humidity) in the
installation location.
When oxygen concentrations fall, the readings may also fall.
Some interference may occur in the presence of highly thermally conductive gases such as CO
2
, Ar,
and He in high concentrations as base gases.
When low vapor pressure gases are detected by aspiration, the gas may condense when aspirated
from warmer areas and detected in cooler areas, or blockages may occur in sintered alloy. Keep
vapor pressure in mind when using the product.
Indication accuracy may degrade in situations in which temperatures fluctuate significantly (±40 °C
or greater) above or below the ambient temperature in which zero calibration or span adjustment
was performed. If this occurs, repeat zero calibration or span adjustment.
There may be fluctuations in readings if the product receives strong shocks from outside or is
subjected to vibrations. If this occurs, perform zero calibration and span adjustment.
The zero point and readings may fluctuate if the sensor comes into contact with high-concentration
gas that is full scale or higher. If this occurs, perform zero calibration and span adjustment.
Readings may fall in environments where gas is continuously present.
The sensor may be hot during replacement. Be careful to avoid burns when replacing the sensor. If
the sensor is hot, turn off the power and wait for it to cool.
There is a Teflon film on the top of the sensors. There may be an impact on the performance of the
sensor if this Teflon film is damaged. Avoid applying excessive pressure or sharp objects to the
Teflon film.
Depending on the conditions and how long the device was off, the reading may oscillate when the
power is turned on. The reading should quickly stabilize at zero.
Regardless of the detection target gas, the warranty period for SGF sensors is 3 years.