7. Maintenance
7-10 Making changes in Environmental setting 2
140 / 191
7-10-23 Caution output setting (CAU OUT)
Set whether to externally output cautions.
Set ON/OFF for the following cautions.
For more information on each caution, refer to ‘10. Troubleshooting’.
E-8X (E-80, E-81, E-82)
E- 9
E- 6
E-94
E-5X (E-50, E-56, E-57, E-58, E-59)
E-15
E-83
E-84
E-98
The default settings are as follows:
EA models, ET models: All cautions are ON
4 - 20 mA models: [E- 6] is OFF, all others are ON
1 In maintenance mode, press the
▲
key or
▼
key to
select [2-10 SETTING2], then press the TEST/SET key.
2 Press
the
▲
key or
▼
key to select [SET-23 CAU OUT],
then press the TEST/SET key.
3 Press
the
▲
key or
▼
key to select the fault code, then
press the TEST/SET key.