Chapter 4 Functions of the Front Panel of EMI
RIGOL
RSA3000E User Guide
4-27
"√" and highlighted, it indicates that auto attenuation within the selected range
is enabled.
13.
RF Preamp
Enables or disables the preamplifier of the selected scan range.
14.
Scan Table
Enables or disables the scan table display. When enabled, the scan table will be
displayed in the user interface.
Scan Sequence
Selects the measurement sequence.
Remarks:
You need to select "Start" under Sweep
Start/Stop menu to start the
selected measurement sequence.
If a scan sequence is currently executed, then the Scan Sequence menu is
grayed out and disabled.
1.
Scan Only
Only performs the pre-scan operation.
2.
Search Only
Only performs the peak search operation.
3.
Scan-Srch-Ms
Performs the pre-scan, peak search, and final measurement.
4.
Scan-Search
Performs the pre-scan and peak search.
5.
Search-Meas
Performs the peak search and final measurement.
6.
(Re)measure
Performs the final measurement.