Chapter 4 Functions of the Front Panel of EMI
4-4 RSA3000E User Guide
Start Freq Mode
Sets the coupling mode of the start frequency to "Manual" or "Auto". When "Auto" is
selected, the start frequency will automatically couple to the start frequency of the
first range among the currently enabled ranges in the scan table.
Stop Freq
Sets the stop frequency of the current frequency channel.
Remarks:
When you modify the stop frequency, the span and midspan frequency will be
changed.
You can use the numeric keys, the knob, or arrow keys to modify this parameter;
you can also use the touch screen to modify the parameter.
Table 4-4 Stop Frequency
Parameter
Remarks
Default
midspan fre span/2
Range
10 Hz to Fmax
Unit
GHz, MHz, kHz, Hz
Knob Step
step = RBW (meter)/2
Left/Right Arrow Key Step
Up/Down Arrow Key Step
step = RBW (meter)*10
Stop Freq Mode
Sets the coupling mode of the stop frequency to "Manual" or "Auto". When "Auto" is
selected, the stop frequency will automatically couple to the stop frequency of the
first range among the currently enabled ranges in the scan table.
Scale Type
Sets the scale type of X-axis to Lin or Log.
Remarks:
The X-axis scale type is only related to data display, and will not affect scan and
trace data.
Modifying the X-axis scale type will neither restart the scan nor affect the
number of scan points.