Chapter 3 Functions of the Front Panel of RTSA
RIGOL
RSA3000E User Guide
3-17
Trace Detector (PvT)
Sets the trace detector for the PvT mode.
Detector Auto (PvT)
Enables or disables the Detector Auto function in PvT mode. By default, Detector
Auto is enabled. If you set the detector type manually, please disable the Detector
Auto function.
Trace Update (PvT)
Enables or disables the trace update in PvT mode.
Trace Display (PvT)
Enables or disables the trace display in PvT mode.
Preset All (PvT)
In PvT mode, turns on Trace 1, and blanks all other traces. This operation does not
affect the trace type, trace detector, and other states.
Clear All (PvT)
Clears all traces in PvT mode. All the trace data will be set to the minimum trace
value (except the case that the trace is in Min Hold state). When the trace is in Min
Hold state, the trace data will be set to the maximum trace value. That is, even if you
turn off the trace update, the trace data will be updated after you perform the "Clear
All" operation
.
Note: The trace settings in PvT and non-PvT modes are independent of each other.
In PvT view, 6 traces are all available for setting, but only one trace can be displayed.