RIGOL
Chapter 5 Functions of the Front Panel of VSA
5-24 RSA3000E User Guide
5.
Symbol
Press or tap the Symbol menu, and the symbols recovered after digital
demodulation is displayed in the selected trace window of the user interface. At
this time, the spectrum analyzer selects the symbols as the trace data.
When this data source is selected, the Trace Format menu is grayed out and
disabled.
6.
Errs Summary
Press or tap the Errs Summary menu, and the corresponding errors summary
is displayed in the selected trace window of the user interface. If you select a
different demodulation type, the displayed error results are different.
When this data source is selected, the Trace Format menu is grayed out and
disabled.
7.
RT Trace
Press or tap the RT Trace menu, and the real-time trace of the signal is
displayed in the selected trace window of the user interface.
When this data source is selected, the Trace Format menu is grayed out and
disabled.
Trace Format
Sets the display format of the selected trace.
1.
Log Mag (dB)
The signal amplitude is converted to be displayed in decibel unit.
2.
Linear Mag
The signal amplitude is displayed in linear unit.
3.
Real (I)
Displays the real part of the signal.
4.
Imaginary (Q)
Displays the imaginary part of the signal.
5.
Wrap Phase
Displays the phase of the data on the Y-axis. It is limited to [-180
º
,180
º
].
6.
Unwrap Phase
Displays the phase of complex data without discontinuities, and the phase is not
limited to [-180
º
,180
º
].
7.
I-Q
Displays the real part of data in the horizontal axis, and displays the imaginary