3-22
RSA3000E User Guide
relationship. You can use the mouse or touch the screen to display one of the
windows in full screen.
The parameter rule in the Density view is consistent with that in the single
Density view. Note that the white real-time trace in the Density view
corresponds to the display trace in the Spectrogram view, but the trace in the
Density view shows the latest data.
The parameter rule in the Spectrogram view is consistent with that displayed in
the single Spectrogram view.
The combined Density and Spectrogram view display the traces, the occurrences
of all the signals within the acquisition interval time, their corresponding
frequency/amplitude and time information.
PvT
Measurement Interface:
Figure 3-6 PvT View
After you select PvT to be the measurement type, the measurement interface is
shown in the figure above. Press Meas Setup to set the relevant parameters.
Remarks:
The PvT measurement provides an analysis for time-domain data. X-axis
represents the acquisition time, and Y-axis represents the power value of the
signal.
In RTSA mode, the acquisition time in PvT measurement can be different from
that in the real-time spectrum measurement, but in the combined display of PvT