User’s Guide for DSA1000 Series
left corner of the display grid, which indicates Span is active.
(2)
Span value could be changed through the numerical keyboard, the knob, or
the directional keys.
(3)
Input 20 and select MHz as its unit using the numerical keyboard, then the
span of the analyzer is set to 20 MHz.
6.
Setup Amplitude
(1)
Press AMPT, Ref Level is in high-bright state, the reference level will be
shown at the top left corner of the display grid, which indicates Reference
Level is active.
(2)
Reference level could be changed through the numerical keyboard, the knob,
or the directional keys.
(3)
If necessary, you can change the reference level using the knob so as to
bring the signal peak value near the top of the grid based on the signal
display.
When all the steps above are finished, a 50 MHz RF curve will be shown on the
analyzer.
7.
Read Result
The value of frequency and amplitude could be read from the Marker. Press
Marker
Select Mkr
1 to activate Marker 1 and set the marker frequency to
“50MHz”, then the frequency and amplitude of where the marker is located would
be shown at the upper right of the display grid.
Figure 1-24 Measuring result