RIGOL
3-8
User’s Guide for DSA1000 Series
Method 2: (Use AM demodulation function)
Procedures:
(1)
Reset the analyzer
-
Press Preset
(2)
Set the parameters
-
Press FREQ
-
Press Center Freq and enter 100 MHz.
-
Press SPAN
-
Press Span and enter 10 MHz.
(3)
Demodulate the signal
-
Press Demod
-
Press Demod and enter the lower menu.
-
Press AM to turn on AM demodulation function.
The frequency can be described by the acuity level of the speaker.
(4)
Set the parameters of AM demodulation function
-
Press Demod Setup and enter the lower menu.
-
Press Speaker to select whether or not to use the speaker under this
function.
-
Press Demod Time and set the desired demodulating time.