RIGOL
User’s Guide for DSA1000 Series
3-13
3.6
Measuring the Time-Domain Power
DSA1030/DSA1030-TG that has installed the Advanced Measurement Kit option can
measure the time-domain power such as the transmitted power in TDMA. In the
example below, we use a generator (such as RIGOL DG3101A) to output a sine signal
with 100 MHz and -10 dBm as the signal under measure.
Procedures:
1.
Device attachment
Connect the output terminal of the generator to the RF INPUT 50Ω terminal on
the front panel of the analyzer.
2.
Measure the time-domain power
(1)
Reset the analyzer
-
Press Preset
(2)
Set the parameters
-
Press FREQ
-
Press Center Freq and enter 100 MHz.
-
Press Meas
-
Press Meas Fctn and enter the lower menu.
-
Press T-Power
Read the measuring result from the measurement frame at the lower left
corner.
(3)
Modify the time power settings
-
Press Meas Setup
-
Press Peak
-
Press Start Line and enter 1 ms.
-
Press Stop Line and enter 9 ms.
3.
Measurement result
The measurement result shown in the lower window is -9.95 dBm, see figure
below.