70 -
|
Page Rigel Medical
62353+
user Manual V1.4
Differential Leakage Measurement
Measuring Range
75μA to 9999μA
Accuracy
±5% of r 5 counts
Measurement / display resolution
1μA
Measurement Type
True RMS
Measuring Device
frequency response characteristics to IEC 60601-1.
Alternative Leakage Measurements:
Test Voltage
250V at mains frequency
Test Current
3.5 mA current limited
Measurement Range
4μA to 9999μA
Measurement Resolution
1μA
Measurement Accuracy
±5% of r 2 counts
Measurement Type
True RMS
Measuring Device
As per IEC 60601-1
Power Measurement
Method
VA rating.
Range
0.1kVA
– 4kVA
Accuracy
±10% + 2 counts
Mains Outlet Test
Input voltage range
0-300V AC
max current
16A
Measures
L
– E, N – E
Accuracy
± 5% of r 2 counts
IEC Mains Lead Test
Test Duration
2s
Test
Continuity of all EB conductor, Insulation & Polarity
General
Mains power
230 VAC ±10%, 50-60Hz +/- 1Hz