39 -
|
Page Rigel Medical
62353+
user Manual V1.4
4 Manual Mode
The Manual mode provides the user with the features of testing a specific
individual test and or test condition for example to aid fault diagnostic
procedures. Manual tests are available from the home screen. Use the Left &
Right Keys to scroll through the various manual tests.
Manual tests can also be chosen, by pressing the
button (F4) and select
Manual Mode from the list. See below.
Do not connect any probe combination to voltages in excess of 30
V AC/DC with respect to earth potential when performing non-
power tests. This can cause damage to the equipment.
Below is a map showing where the individual tests can be selected
in the Manual Mode menu.
Manual Mode >
Earth Bond
Insulation >
Insulation EUT
Insulation AP
Insulation AP to mains
IEC 62353 >
Equipment Leakage >
Direct
Alternative
Differential
AP Leakage Test >
Direct
Alternative
IEC lead test