40 -
|
Page Rigel Medical
62353+
user Manual V1.4
For example; To select an IEC 62353 - Equipment Leakage - Direct Method;
Use the Up, Down, Left and Right keys to navigate through the menu and press
the
button (F4) to confirm.
The following manual mode screen appears;
The centre of the analogue bargraph indicates the limit value (in this
case 500µA)
Warning, during many of the leakage tests the PROTECTIVE EARTH
CONDUCTOR to the appliance under test is interrupted. For this reason
additional means of protection are required in order to comply with the
relevant safety standards.