41 -
|
Page Rigel Medical
62353+
user Manual V1.4
The screen shows the current measuring condition, ie;
Equipment Leakage (Direct Method)
2 second test duration
limit is [500µA]
Single Fault Condition (normal polarity
– open earth)*
Class I and enclosure probe symbol
* The IEC 62353 specifies that the Direct Method measurement is done using an
interrupted earth. The
button (F2) will only reverse the power supply.
All manual tests can be modified to meet user requirements. To change the
settings in the manual mode, simply press the
button (F3).
Help is available by pressing the
button (F1) in the form of a schematic
representation of the test.
Depending of the type of test, the F2 key is either used to select the test class
(Class I or II) or to set the single fault conditions. In the latter case, the test class
can be set within the
menu button (F3).
4.1.1 Earth Continuity
This test is applicable to Class I equipment only.
Press the F1
button to obtain the schematic representation of this test:
If an earth bond lead other than that supplied with the Rigel 62353 Plus is to be used
then any resistance associated with this lead can be zeroed out by connecting the
supplied earth bond lead to the EUT earth socket on the Rigel 62353 Plus and