32 -
|
Page Rigel Medical
62353+
user Manual V1.4
Visual Test
This is the first test in the test sequence (when activated, not part of default test
sequences);
Press the corresponding appropriate button to select the outcome of the visual test.
Custom Visual tests can be created. See 2.1.6 for more information.
If one or more Visual tests result in a failure, a Failure menu will
appear.
Provided the test has Passed, the next test will be the Earth Continuity
Attention! Ensure the earth bond test lead is connected to the black socket when
testing between probe and EUT earth.
Press the Calc button (F2) to calculate the pass fail limit based on the type of cable
(mm²) and cable length.
Pressing the
button (F3) will bring up the failure menu (see 3.5.1)