52
−
ME12058B
Operation Section
9.7 XRR/XRD Measurement
Troubles
Causes and remedies
The measurement profile
intensity is extremely high.
Check to make sure that the slit setting in the recipe.
The measurement profile
intensity is extremely low.
Check to make sure that the sample alignment is conducted
properly.
Check to make sure that the surface/interface roughness is not
excessive.
If the surface/interface roughness is excessive (several nm or
larger), the reflection intensity decreases rapidly, disallowing
the observation of oscillations according to film thickness.
The critical angle cannot be
fitted.
Check to make sure that the curvature of the sample is not too
large.
If the curvature of the sample is too large, conduct the sample
alignment using the auto focus.
Alignment of the critical angle of Si can be achieved by
setting the 2
θ
offset (2
θ
0).
Analysis cannot be performed
properly.
Refer to the analysis know-how section in the analysis
software manual.
NOTE:
Contact the maintenance personnel for
checking the slit condition in the recipe.
NOTE:
Contact the maintenance personnel for
checking the alignment condition in the
recipe.
NOTE:
Contact the maintenance personnel for
checking the recipe.
Summary of Contents for MFM65S2RFFFLFL
Page 19: ...14 ME12058B Operation Section Goniometer Figure Fig 3 1 2 Gonimeter...
Page 37: ...32 ME12058B Operation Section Fig 5 2 5 Power Switch 5 6...
Page 67: ...62 ME12058B Operation Section Fig A3 1 2 Power Circuit Diagram 2...
Page 69: ...64 ME12058B Operation Section A4 2 MSDS of Beryllium...
Page 70: ...65 ME12058B Operation Section...
Page 71: ...66 ME12058B Operation Section...
Page 72: ...67 ME12058B Operation Section...
Page 73: ...68 ME12058B Operation Section...
Page 74: ...69 ME12058B Operation Section...
Page 75: ...70 ME12058B Operation Section...
Page 76: ...71 ME12058B Operation Section...
Page 77: ...72 ME12058B Operation Section...
Page 78: ...73 ME12058B Operation Section...
Page 79: ...74 ME12058B Operation Section...
Page 80: ...75 ME12058B Operation Section A4 3 MSDS of Lead...
Page 81: ...76 ME12058B Operation Section A4 4 MSDS of Lithium...
Page 82: ...77 ME12058B Operation Section...
Page 83: ...78 ME12058B Operation Section...
Page 84: ...79 ME12058B Operation Section A4 5 MSDS of Mercury...
Page 85: ...80 ME12058B Operation Section...
Page 86: ...81 ME12058B Operation Section...
Page 87: ...82 ME12058B Operation Section...
Page 88: ...83 ME12058B Operation Section...
Page 89: ...84 ME12058B Operation Section...
Page 90: ...85 ME12058B Operation Section A4 6 MSDS of Mechanical Lubricating Grease...
Page 91: ...86 ME12058B Operation Section...
Page 92: ...87 ME12058B Operation Section...
Page 93: ...88 ME12058B Operation Section...
Page 94: ...89 ME12058B Operation Section...
Page 95: ...90 ME12058B Operation Section...
Page 96: ...91 ME12058B Operation Section...
Page 97: ...92 ME12058B Operation Section A4 7 MSDS of Ethanol...
Page 98: ...93 ME12058B Operation Section...
Page 99: ...94 ME12058B Operation Section...
Page 100: ...95 ME12058B Operation Section...
Page 101: ...96 ME12058B Operation Section...