46
−
ME12058B
Operation Section
9.2
Startup
Troubles
Causes and remedies
The goniometer cannot be
initialized.
Check to make sure that the side doors of the main module or
EFEM are not open. (See “2.3 Interlock List”.)
Check to make sure that the load port panels or rear panels
are not open.
Opening any of these panels or doors turns off the power of
the goniometer. Be sure to close the side doors and panels.
If the cause of the problem is unknown, shut down the PC,
turn off the main breaker lock-out handle of the equipment,
then restart the equipment.
(Refer to “2.4 Lock-out/Tag-out”.)
The sample transfer system
cannot be initialized.
Check to make sure that the side doors of the main module or
EFEM are not open. (See “2.3 Interlock List”.)
Check to make sure that the load port panels or rear panels
are not open.
Opening any of these panels or doors turns off the power of
the sample feed system. Be sure to close the side doors and
panels.
Check to make sure that the manual robot controller (inside
the EFEM) is not set to the manual mode.
Set the manual robot controller to the remote mode ([RMT] is
indicated).
Check to make sure that CDA and vacuum are supplied.
Make sure of utility connections.
If the cause of the problem is unknown, shut down the PC,
turn off the main breaker lockout handle of the equipment,
then restart the equipment.
(Refer to “2.4 Lock-out/Tag-out”.)
CAUTION
When checking the manual robot controller,
contact the maintenance personnel.
NOTE:
When checking the CDA,
contact the maintenance personnel.
Summary of Contents for MFM65S2RFFFLFL
Page 19: ...14 ME12058B Operation Section Goniometer Figure Fig 3 1 2 Gonimeter...
Page 37: ...32 ME12058B Operation Section Fig 5 2 5 Power Switch 5 6...
Page 67: ...62 ME12058B Operation Section Fig A3 1 2 Power Circuit Diagram 2...
Page 69: ...64 ME12058B Operation Section A4 2 MSDS of Beryllium...
Page 70: ...65 ME12058B Operation Section...
Page 71: ...66 ME12058B Operation Section...
Page 72: ...67 ME12058B Operation Section...
Page 73: ...68 ME12058B Operation Section...
Page 74: ...69 ME12058B Operation Section...
Page 75: ...70 ME12058B Operation Section...
Page 76: ...71 ME12058B Operation Section...
Page 77: ...72 ME12058B Operation Section...
Page 78: ...73 ME12058B Operation Section...
Page 79: ...74 ME12058B Operation Section...
Page 80: ...75 ME12058B Operation Section A4 3 MSDS of Lead...
Page 81: ...76 ME12058B Operation Section A4 4 MSDS of Lithium...
Page 82: ...77 ME12058B Operation Section...
Page 83: ...78 ME12058B Operation Section...
Page 84: ...79 ME12058B Operation Section A4 5 MSDS of Mercury...
Page 85: ...80 ME12058B Operation Section...
Page 86: ...81 ME12058B Operation Section...
Page 87: ...82 ME12058B Operation Section...
Page 88: ...83 ME12058B Operation Section...
Page 89: ...84 ME12058B Operation Section...
Page 90: ...85 ME12058B Operation Section A4 6 MSDS of Mechanical Lubricating Grease...
Page 91: ...86 ME12058B Operation Section...
Page 92: ...87 ME12058B Operation Section...
Page 93: ...88 ME12058B Operation Section...
Page 94: ...89 ME12058B Operation Section...
Page 95: ...90 ME12058B Operation Section...
Page 96: ...91 ME12058B Operation Section...
Page 97: ...92 ME12058B Operation Section A4 7 MSDS of Ethanol...
Page 98: ...93 ME12058B Operation Section...
Page 99: ...94 ME12058B Operation Section...
Page 100: ...95 ME12058B Operation Section...
Page 101: ...96 ME12058B Operation Section...