5010e_513.doc / 22.01.10
7.2.4 Pump
menu
7.2.4 Pump
menu
With these functions the pump and the bubble detector can be checked and adjusted. The check and adjust-
ment is possible only with inserted flow-through cuvette adaptor.
With these functions the pump and the bubble detector can be checked and adjusted. The check and adjust-
ment is possible only with inserted flow-through cuvette adaptor.
The menu offers following functions:
•
Activation and deactivation of pump
•
Calibration of pump volume
•
Activation and deactivation of bubble
detector and of the volume optimized
mode.
•
Media test: check and adjustment of
bubble detector.
7.2.4.1 Pump ON / OFF
The current status of the pump is indicated in
the first line by OFF or ON.
Change setting by [SELECT]
Save setting permanently by [OK]
Save setting temporarily up to next switch-off
of the device by [EXIT]
7.2.4.2 Pump calibration
The pump calibration can be executed with or
without bubble detector. So set bubble detec-
tor ON or OFF.
Empty aspiration tube by [WASH].
Put 1000 µl distilled water into a sample cup.
Dip aspiration tube into the sample cup up to
the base.
The volume of 1000 µl is sucked in by
[Sipping lever P].
Behavior
with
bubble detector:
After stoppage of the pump values for air and
water are indicated which correspond to the
delivered volumes.
Save setting permanently by [OK]
If the calibration is not possible,
although the pump tube is connected and
1000 µl were sucked in, execute a media test
(see 7.2.4.4). Afterwards the pump calibration
can be repeated.
U T I L I T I E S S U B M E N U
P U M P O N / O F F
P U M P C A L I B R A T I O N
P U M P S E T T I N G S
M E D I A T E S T E X I T
3 7 . 0 5 C 0 9 / 0 8 / 0 9 1 1 : 3 9
P U M P O N
E X I T O K S E L E C T
P U M P C A L I B R A T I O N
E X I T W A S H
P U M P C A L I B R A T I O N
A I R 1 5 9
W A T E R 1 0 9 6
M S / 5 0 µ l
E X I T O K W A S H
ROBERT RIELE GmbH & Co KG
Page 47
7- UTILITY PROGRAMS
Y PROGRAMS