MAINTENANCE
8 MAINTENANCE
This chapter provides necessary information concerning general maintenance by the user.
If any faults should occur which cannot be remedied, then service should be contacted. Repairs at the
device may be carried out only by authorized specialist staff. Through improper repairs the warranty extin-
guishes, and the user can be heavily jeopardized.
8.1
CLEANING INSTRUCTION
Liquid waste is potentially biologically hazardous. Always wear gloves if handling those materials. Do not
touch parts of the device other than those specified. Consult the laboratory protocol for handling biohazardous
materials.
Take care that no liquid enters the device! There is no protection against penetrating of liquids
(Code IP X0). Check from time to time that the tubing and the connections are leakproof.
The
flow-through system
of the Photometer
5010
has to be washed with dist. water regularly before and after
measuring 2 – 3 times. Depending on sample material and reagent, and
always
at the end of a working day, the
Photometer
5010
has to be cleaned additionally with a phosphate-free detergent, e.g. with approx. 5ml of
Biogent-A (REF 5010-024), and afterwards rinsed with 5ml distilled water.
Persistent residues are to be removed with a combined alkaline-acid treatment. The following procedure is
advised step by step:
1. 5 x wash with NaOH 1 N
2. 5 x rinse with dist. water
3. 5 x wash with HCl 0.5 N
4. 5 x rinse with dist. water
If the system is contaminated severely, it may alternatively be cleaned with hypochlorite. The following proce-
dure is advised step by step:
1. 5 x rinse with dist. water
2. 2 – 3 x wash with hypochlorite (1:20 diluted solution) or ISE Cleaning Solution,
undiluted, let it possibly affect up to 20 minutes
3. 5-10 x rinse with dist. water
For device cleaning and surface decontamination purposes use commercial decontaminating solution which are
usually available in clinical chemistry laboratories like Mikrozid
®
AF Liquid, Bacillol
®
plus, 3 % Kohrsolin
®
or
similar solutions. Switch off the device and disconnect it from the mains voltage. Then clean the device with soft
cloth and decontaminating solution.
Empty the drain tank at the end of daily measurement, or whenever filled.
8.2
CALIBRATING MEASURING SYSTEM
At doubtful measurement results a dark adjustment has to be carried out corresponding to chapter 7.2.1.
8.3
ADJUSTMENT OF BUBBLE DETECTOR
See chapter 7.2.4 - Pump
8.4
CALIBRATION OF PERISTALTIC PUMP
See chapter 7.2.4 - Pump
8.5
REPLACEMENT OF RIBBON
Open printer cover. Lift used ribbon by pressing “PUSH” on the left. Take ribbon cassette out of the printer.
Take new ribbon cassette out of the wrapping. Turn small wheel on the left in the direction marked by an arrow
till the inked ribbon is tightened.
Put in the ribbon cassette. Pay attention that the paper is to be taken through inked ribbon and cassette. Take
printer paper through the crack of the printer cover and close the printer cover.