Calculation procedure .....................................................CP 14
Characteristic ........................................................ C / F / Delta
Method ...................................................Difference with Factor
Calculation formula ..................... C = F
∗
(
∆
A
S2-Sb2
-
∆
A
S1-Sb1
)
Factor............................................................... given / entering
Sample blank .......................................................with / without
Time / Delta T1..................................entering (0, 10s to 255s)
Measure time T2 .....................................entering (0 to 1800s)
Delay T3..................................................entering (0 to 1800s)
Reagent time #2.......................................entering (0 to 1800s)
Reagent time #3.......................................entering (0 to 1800s)
PHOTOMETER 5010 #7000
V5.13a 02/09/09 D
LAB.: RIELE GMBH+CO KG
DATE: 09/08/09
TIME: 07:16:31
METHOD 14: C/F DELTA
PROGRAM: 14
FACTOR: 1.000
WAVELENGTH: 405nm
TEMPERATURE: 37C
TIME/DELTA: 12s
MEASURE TIME: 100s
DELAY: 10s
REAGENT 2: 40s
REAGENT 3: 60s
UNIT: U/l
- - - - - - - - - - - -
MEASURE BLANK
NO. Sb[A] S[A]E1
1 0.083 0.411
2 0.110 0.382
3 0.146 0.492
NO. Sb[A] S[A]E2
1 0.091 1.090
2 0.140 0.991
3 0.200 1.165
NO. RESULT
1 0.671
2 0.578
3 0.619
- - - - - - - - - - - -
NO. Sb[A] S[A]E1
1 0.000 1.012
2 0.000 1.138
3 0.000 1.076
NO. Sb[A] S[A]E2
1 0.000 1.458
2 0.000 1.530
3 0.000 1.384
NO. RESULT
1 0.446
2 0.392
3 0.307
Start method selection in the main menu.
See chapter:
4.1 - Measurement with programmed methods
4.2 - Measurement with basic methods
In the case of activated printer the print out of the method data
follows.
The measuring window is shown.
Method procedure with sample blank:
Æ
Insert / measure zero solution
Æ
Measure all samples E1 (maximum 25)
Æ
Change to measuring E2 by [E1/E2]
Æ
Measure all samples E2 (maximum 25)
Results based on the differences of the measured samples
Æ
Show the results by [MODE] [MODE] [DETAIL]
Method procedure without sample blank:
Æ
Measure all samples E1 (maximum 25)
Æ
Change to measuring E2 by [E1/E2]
Æ
Measure all samples E2 (maximum 25)
Results based on the differences of the measured samples
Æ
Show the results by [MODE] [MODE] [DETAIL]
ROBERT RIELE GmbH & Co KG
Page 38
5 - CALCULATION PROCEDURES