5.3 EDP Command Calibration
Use the following instructions to calibrate the 680HE using EDP commands. For information on the EDP commands of the
680HE, see
.
The indicator must respond with OK after each step or the calibration procedure must be done again.
1. Place indicator into setup mode via the setup jumper (
).
2. For a standard calibration, remove all weight from scale (except hooks or chains which are needed to attach weights).
3. Send the command
SC.WZERO#s
to perform a standard calibration of the zero point.
• Send
SC.TEMPZERO#s
to perform a temporary zero calibration
• Send
SC.LASTZERO#s
to perform a last zero calibration
4. Apply the span calibration weight to the scale.
5. Send the command
SC.WVAL#s=xxxxx
, where
xxxxx
is the exact value of the span calibration weight applied to the
scale.
6. Send the command
SC.WSPAN#s
to calibrate the span point. Continue on to
to calibrate additional
linearization points, or proceed to
.
7. Apply weight equal to the first linearization point to the scale.
8. Send the command
SC.WLINVn#s=xxxxx
, where
n
is the linearization point number (1-4) and
xxxxx
is the exact
value of the weight applied.
9. Send the command
SC.WLIN.Cn#s
to calibrate the linearization point.
for up to four total linearization points.
11. If hooks or chains were used to attach the weights, remove all weight, including the hooks and chains, and send the
command
SC.REZERO#s
to remove the zero offset.
12. Send the command
KSAVEEXIT
to return to weigh mode.
Note