680HE Synergy Indicator
24
Visit our website
Ovrloa
Overload – Determines the point at which the display blanks and the overload error message displays (^^^^^^);
Maximum legal value varies depending on local regulations;
Settings:
FS+2%
(default), FS+1D, FS+9D, FS
Smprat
Sample Rate – Selects measurement rate, in samples per second, of the analog-to-digital converter; Lower sample rate values
provide greater signal noise immunity;
Settings: 6.25HZ, 7.5HZ, 12.5HZ, 15HZ, 25HZ,
30HZ
(default), 50HZ, 60HZ, 100HZ, 120HZ
fltchn
Filter Chain Type – Sets the filter type to be used;
Settings:
AVGONL
(default) – Digital Rolling Average Filter (
); Uses DIGFL1-3, DFSENS and DFTHRH
ADPONL – Adaptive Filter (
); Uses ADSENS and ADTHRH
DMPONL – Damping Filter (
); Uses DMPVL
RAW – No filtering
digfl1-3
Digital Filters – Sets the digital filtering rate used to reduce the effects of environmental influences from the immediate area of
the scale; Settings indicate the number of A/D conversions per update which are averaged to obtain the displayed reading; a
higher number gives a more accurate display by minimizing the effect of a few noisy readings, but slows down the response time
of the indicator;
Settings: 1, 2,
4
(default), 8, 16, 32, 64, 128, 256
Dfsens
Digital Filter Sensitivity – Specifies the number of consecutive A/D readings which fall outside the Filter Threshold before filtering
is suspended;
Settings:
2OUT
(default), 4OUT, 8OUT, 16OUT, 32OUT, 64OUT, 128OUT
Dfthrh
Digital Filter Threshold – Sets a threshold value, in display divisions; when a number of consecutive A/D readings (Digital Filter
Sensitivity) falls outside of this threshold value (when compared to the output of the filter), filtering is suspended and the A/D
value is sent straight through the filter; Filtering is not suspended if the threshold is set to NONE;
Settings:
NONE
(default), 2D,
5D, 10D, 20D, 50D, 100D, 200D, 250D
Adsens
Adaptive Filter Sensitivity – Controls the stability and response time of the scale;
Settings:
LIGHT
(default)
– Fastest response to small weight changes, but less stable
MEDIUM
– Has a quicker response time than heavy, but more stable than light
HEAVY
– Results in an output which is more stable but settles slowly; small changes in weight data (a few grads) on the
scale base are not seen quickly
Adthrh
Adaptive Filter Threshold – Sets the adaptive filter weight threshold value (in display divisions); a weight change exceeding the
threshold resets the filtered values; must be set above the noise disturbances in the system (if set to zero, the filter is disabled);
Enter value: 0–2000,
10
(default)
Dmpvl
Damping Value – Sets the damping time constant (in 0.1 sec intervals);
Enter value: 0–2560,
10
(default)
Rtltrp
RattleTrap – Enables RattleTrap filtering; Effective at eliminating vibration effects, environmental influences and mechanical
interference from nearby machinery, may increase response time over standard digital filtering;
Settings:
OFF
(default), ON
Tarefn
Tare Function – Enables or disables push-button and keyed tare;
Settings:
BOTH
(default)
– Both push-button and keyed tare are enabled
NOTARE
– No tare allowed (gross mode only)
PBTARE
– Push-button tares enabled
KEYED
– Keyed tare enabled
Sstime
Standstill Time – Specifies the length of time the scale must be out of motion, before the scale is considered to be at standstill (in
0.1 sec intervals);
Enter value: 0–600,
10
(default)
sense
Sense – Specifies the type of load cell cable connected to the J1 connector (
);
Settings:
4-WIRE
(default), 6-WIRE
Menu
Description
Table 4-4. Setup – Configuration Menu Descriptions (Continued)