Calibration
The 680HE can be calibrated using the front panel and EDP commands. The following sections describe the procedures
required for both of these calibration methods.
The 680HE requires a WZERO and WSPAN points to be calibrated. The linear calibration points are optional; they
must fall between zero and span, but must not duplicate zero or span.
Figure 5-1. Calibration Menu
5.1 Front Panel Calibration
5.1.1 Span Calibration
Use the following steps to perform a standard span calibration on a connected scale.
1. Access the setup menu via the setup jumper (
).
Config
displays.
2. Press
twice.
Calibr
displays.
3. Press
.
wzero
displays.
See
if the application requires a rezero, last zero or temporary zero.
4. Ensure there is no weight on the scale.
5. Press
to perform a zero calibration.
ok
displays.
6. Press
.
zrocnt
for more information on
zrocnt
.
7. Press
.
wval
displays.
8. Press
. The current test weight value displays.
9. Press
and use the numeric keypad to enter a new value, if necessary.
10. Press
to accept value.
wSpan
displays.
11. Place the specified amount of test weight on the scale.
12. Press
to perform a span calibration.
ok
displays.
13. Press
.
spncnt
displays. See
zpncnt
.
Span calibration is complete. To continue with a linear calibration, see
before returning to
weigh mode.
14. Press
to return to weigh mode.
Note
Setup
Calibr
zrocnt
WZERO
WVAL
spncnt
WSPAN
Point1-4
WLIN
REZERO
Wlinv1-4
wlinc1-4
wlinf1-4
lstzro
tmpzro
Note
Note