38
IQ plus 710 Installation Manual
5.1.2
Reporting Commands
Reporting commands (see Table 5-2) send specific
information to the EDP port. These commands can be
used in both setup mode and normal mode.
5.1.3
The RESETCONFIGURATION Command
The RESETCONFIGURATION command can be
used to restore all configuration parameters to their
default values. Before issuing this command, the
indicator must be placed in test mode (press and hold
setup switch for approximately three seconds).
This command is equivalent to using the DEFAULT
function in TEST mode. See Section 10.8 on page 69
for more information about test mode.
NOTE:
All load
c e l l c a l i b r a t i o n s e t t i n g s a r e l o s t w h e n t h e
RESETCONFIGURATION command is run.
5.1.4
Parameter Setting Commands
Parameter setting commands allow you to display or
change the current value for a particular configuration
parameter (Tables 5-3 through 5-12).
Current configuration parameter settings can be
displayed in either setup mode or normal mode using
the following syntax:
command
<ENTER>
Most parameter values can be changed in setup mode
only; setpoint parameters listed in Table 5-8 on
page 41 can be changed when in normal weighing
mode.
Use the following command syntax when changing
parameter values:
command
=
value
<ENTER>
where
value
is either a number or a parameter value.
Use no spaces before or after the equal (=) sign. If you
type an incorrect command, the display reads
??
.
For example, to set the motion band parameter to 5,
type the following:
MOTBAND
=
5D
<ENTER>
Command
Function
DUMPALL
List all parameter values
VERSION
Write IQ plus 710 software version
P
Write current displayed weight with units
identifier. See Section 10.2 on page 63 for
more information.
S
Write one frame of stream format
Table 5-2. EDP Reporting Commands
Command
Description
Values
GRADS
Graduations
1–100000
ZTRKBND
Zero track band
OFF, 0.5D, 1D, 3D
ZRANGE
Zero range
1.9%, 100%
MOTBAND
Motion band
1D, 2D, 3D, 5D, 10D, 20D, OFF
OVRLOAD
Overload
FS+2%, FS+1D, FS+9D, FS
DIGFLTR
Digital filtering
1, 2, 4, 8, 16, 32, 64, 128, 256, 1RT, 2RT, 4RT, 8RT, 16RT, 32RT,
64RT, 128RT, 256RT
DFSENS
Digital filter cutout sensitivity
2OUT, 4OUT, 8OT, 16OUT, 32OUT, 64OUT, 128OUT
DFTHRH
Digital filter cutout threshold
NONE, 2DD, 5DD, 10DD, 20DD, 50DD, 100DD, 200DD, 250DD
PWRUPMD
Power up mode
GO, DELAY
TAREFN
Tare function
BOTH, NOTARE, PBTARE, KEYED
Table 5-3. CONFIG EDP Commands
Command
Description
Values
PRI.DECPNT
Primary
units
decimal
position
8.888888, 88.88888, 888.8888, 8888.888, 88888.88, 888888.8,
8888888, 8888880, 8888800
PRI.DSPDIV
Primary units display divisions
1D, 2D, 5D
PRI.UNITS
Primary units
LB, KG, G, OZ, TN, T, GN, TROYOZ, TROYLB, LT, NONE
Table 5-4. FORMAT EDP Commands