KrosFlo® TFDF™ Lab System
User Guide
70
repligen.com IF.UG.023 R1
System specifications
Table 9. System output
Description
Specifications
Feed / recirculating pump type
Magnetic levitating pump
Feed / recirculating pump capacity
0 - 11000 RPM, 0 - 10 LPM @ 0.0 bar,
0.0 - 21.8 psi (1.5 bar)
Diafiltration and permeate pumps type
Peristaltic
Diafiltration and permeate pumps capacity
0.1 - 100 RPM (0.01 RPM Resolution)
Maximum 340 ml/min (4.8 mm ID and 1.6 mm
thickness)
0.0002 - 35 ml/min/channel
3 channel, 8 rollers
14.5 psi (1.0 bar) max differential pressure
Display
Automation Direct
12” LCD touchscreen
Retentate flow meter
Ultrasonic clamp-on flow meter
0 - 8000 ml/min, 2% accuracy (±16 ml/min)
Calibrated for #15 PharmaPure® tubing
Recommended process volume
1 - 50 L
Number of pressure sensors supported
5
Pressure sensor range
-14 - 30 psi (-1 - 2 bar)
TFDF™ Filter surface area supported
2 - 150 cm
2
Number of scales supported
2
Table 10. System input
Description
Specifications
Power requirements
120 VAC, 10 A
230 VAC, 5 A, 50/60 Hz
Table 11. System construction
Description
Weight
Controller weight
36 lbs (16.2 kg)
Controller dimensions
16 x 13 x 21 in (40 x 33 x 53 cm)
Pump Station weight
16 lbs (7.3 kg)
Pump Station dimensions
11 x 11 x 19 in (min)/39 in (max)
(28 x 28 x 48/99 cm)
Controller type
PLC
Controller and Pump Station rating
IP20
Enclosure material of construction
Delrin and powder coated/anodized aluminum