Figure 22. Admin screen
Changing the default tubing calibration
factors
The calibration factor converts pump rotations to a volumetric flow rate. Default values are included
with the system. Values specific for your unit and tubing require measuring volume transfer over a
period of time with a set pump rpm. Changing default values is done at the Admin level only. To
change a tubing calibration factor setting, select one of the blue fields and enter a new default value.
Setting the calibration factor here sets the default for operation.
Users can restore tubing calibration factors to default values using the
Reset Tubing Calibration
button in the System Settings screen. For more information, see the
Changing the system serial number
Select
Serial #
and enter the new system serial number. Once the number is updated, the new value
will display under Serial #.