Manual Mode
In the Overview screen in Manual Mode, the pumps can be started manually by pressing the red
Start
button. Alternatively, the feed (P-01) and permeate (P-03) pumps can be put into an individual
automation loop where data from either the flow meter modulates pump speed. Access this feature
by toggling the
Manual
button to
Auto
. Only the individual control loop for that pump is activated
and the pump will run at the setpoint entered based on the feedback from the flow meter (FL-01)
for the feed pump (P-01) control or scale (WE-01) for the Aux/Diafiltration pump (P-03) control.
Figure 45. Manual Mode Overview screen
Any combination of running auxiliary components is possible in Manual Mode, there are no set
points to automate:
•
Start/stop main pump (P-01)
•
Start/stop auxiliary pumps (P-02 and P-03)
•
Change direction of auxiliary pumps (P-02 and P-03)
•
Tare scales (WE-01 and WE-02)
•
Tare pressure sensors (PE-01, PE-02, PE-03, PE-04, PE-05, and PE-06)
•
Tare Flow meter (FL-01 and FL-02)
Description of Terms:
•
Perm Flow:
Flow of permeate calculated by rpm of permeate pump
•
VT:
Volumetric Throughput - Total permeate mass/volume divided by the surface area of the
filter
•
Perm Total:
Totalized Permeate volume calculated from pump rpm