Care and maintenance
Optics
Cleaning of the optics should be a last resort
To maintain system performance, the XL optics must be kept clean by
following good handling practice:
•
Do not touch the optical surfaces�
•
Minimise use in contaminated environments�
•
Store securely when not in use�
Cleaning recommendations
•
Only use approved solvents for cleaning the optics: methylated spirit
(preferred) or optical grade isopropyl alcohol (IPA)�
•
Wipe only with non-abrasive lens tissue or lint-free cloth wrapped around
a cotton bud (do not use a cotton bud directly on the optic, as this may
increase debris)�
•
Clean the optics using a gentle action� Never use a scrubbing action,
as this might damage the coatings�
Failure to follow these recommendations may lead to damage to the coatings
and glass elements of the optics�
Care of small linear optics kit
If contamination of the optical surfaces is suspected, undo the four cap
screws holding the optics cartridge within the housing� Carefully invert the
unit to separate it from the housing, and clean the optical surfaces using the
recommended method�
When reassembling, ensure that the optics carrier is aligned against
one of the long edges of the housing before fully tightening the cap screw�
This ensures proper alignment of the optics within the housing�