47
XL80 laser system
XL laser calibration
Calibration certificate
Each XL laser is delivered with a calibration certificate. This demonstrates that
the system has been calibrated at the Renishaw factory against reference
systems with traceability to National Standards� It is proof of the equipment’s
performance as tested before delivery� Visit the
webpage for more information�
The certificate is a valuable document and may be required to satisfy your
own or your customer’s quality assurance requirements�
Duplicate documents can be supplied at an additional charge�
Certificate content
Each certificate is unique and is identified by a certificate number. All XL laser
certificates provide the following key information:
•
Serial number of calibrated XL laser
•
Specific test results and graph
•
Statement of accuracy
•
Traceability data (calibration details, see below)
•
Test conditions and methodology
The first page gives graphical representation of test results and specification
limits, as well as tabulated details of the specific test results and uncertainty
of measurement� User can visually verify whether or not the unit meets
published specifications and/or your own requirements. To enable traceability,
details of the test equipment used are given� The date of testing and the date
of printing the certificate are separately noted and the results are signed by a
Renishaw-authorised engineer�
The second page gives generic details of the test procedure, test
environment, and applicable standards, in accordance with ISO 17025�
EXAM
PLE