27
If the machine malfunctions or the following
messages appear, the engineers may ask you to
download the error records for them to diagnose
the machine in advance.
Please select Alarm History.
Save files
Insert the USB Flash Drive and then press the
Save button to store the current data after the
USB icon shows in the status bar. Please take out
the USB Flash Drive and then you can perform
other operations after the icon of USB Flash Drive
disappears.
━ Alarm History
Prev
Maintenance
Update Program
Vender
Piercing Needles
System Status
Optic Test
Alarm History
Back
Clear
Save to USB
Alarm Message
Back
Date
Time
Error Messages
Code
Alarm Message
Back
0000001
Error Code
Please refer to user manual or
contact your local engineer.
Error
Please save the data in the USB Flash Drive
and transfer to computer, or the data in
the USB Flash Drive may be overwritten
later.
File name: AlarmHistory.cvs
━ UV sterilization
Please select the UV button.
Please select the sterilization time.
Maintenance
Reject Tip
Piston Test
UV
Heater Test
Replace O-Ring
Motor Test
Next
Back
30
Select Min :
UV Sterilization
15 min
30 min
60 min
Back
Start
Warning!
The UV lamp is dangerous.
Please do not look straight at
the inner of the machine when
operating the machine.
Summary of Contents for MagCore HF16 Plus
Page 1: ...MagCore Super HF16Plus EngineerInTraining manual ...
Page 34: ...33 Disassemble thebackservicelid Disassemble back service lid ...
Page 35: ...34 RBCBioscience Disassemblethe topservicelid Po ...
Page 63: ...62 RBCBioscience Maintenance Electricalcontrolandengineeringsoftware ...
Page 66: ...65 Mainboardconfiguration ...
Page 75: ...74 RBCBioscience PositionTeachingTools 200µlSPTip toolA toolB toolC PipetteTip ...
Page 87: ...86 RBCBioscience ...
Page 93: ...92 RBCBioscience ...
Page 94: ......
Page 95: ......