Replace the Y-axis belt
Open the back service lid.
1. Install the Y-axis fixing plate on the machine and screw up the two screws (Phillips
screwdriver).
2. Screw off the four screws of the metal sheet of the Y-axis motor (3mm Hex wrench).
3. Push the metal sheet of the motor forward.
4. Replace a new timing belt.
5. Install the Y-axis motor orderly on the machine.
6. Note that the belt should be a little tight when reinstalling it.
7. Disassemble the Y-axis fixing plate and screw off the two screws (Phillips
screwdriver).